Skip to main content

LiquiMAX G6PDH (Quantitative/ Kinetic)

Quantitative/ Kinetic

INTENDED USE: LiquiMAX G6PDH is an in-vitro diagnostic kit for the quantitative determination of Glucose-6-phosphate dehydrogenase in human blood on photometric systems.

Product Fearures:

  • Liquid Stable, Ready to use Three Reagents (R1 + R2 + Lysing Reagent).
  • NADP Analogues are used for better stability
  • 4 Minutes increasing Kinetic Reaction (60 Sec Delay+ 180 Sec Measuring) (For testing the Random Clinical Samples)
  • 10 Minutes increasing Kinetic Reaction(300 Sec Delay+ 300 Sec Measuring) (For the confirmation of deficient samples)
  • Linearity: 25 (U/g Hb).
  • Measuring Wavelength 340 nm
  • Kinetic Factor : 4839 (U/g Hb) at 340 nm at 37° C.
  • Whole blood is the specimen
  • Available as multipurpose reagents.

Categories:

Ref./Cat. No.
AVG6P1-10 | AVG6P1-20
Pack Size
10 ml | 20 ml
Presentation
Three Reagents | Three Reagents
  • Reaction type : Kinetic
  • Reaction Direction : Increasing
  • Wavelength : 340 nm
  • Flow cell temp : 37°C
  • Zero setting with : Distilled water
  • Delay time : 60 Sec
  • Measuring/Read Time : 60 Sec
  • No. of readings : 4
  • Reagent volume : 0.5 ml (Buffer) + 0.5 ml (Substrate)
  • Sample Volume : 0.5 ml (Hemolysate)
  • Factor : 4839 (When Hemoglobin is measured) 48390 (When RBC Count is measured)
  • 12 Linearity : Up to (22.5 U/g Hb) or (628 U/10 RBC’s)

Related Products