Skip to main content

LiquiMAX CK-MB

(IFCC Method)

INDENTED USE: Kit is use for the quantitative determination of CK-MB Iso Enzyme in Human Serum and Plasma.

Product Fearures:

  • Liquid Stable, Ready to use Two Reagents. (R1= Buffer, R2=Substrate)
  • Kinetic Factor 8655
  • NADP analogues are used for better stability.
  • Linearity 1000 IU/L.
  • 8 Minutes increasing Kinetic Reaction
  • Measuring wavelength 340 nm.
  • Serum and Plasma are the specimens
  • Available as multipurpose reagents and dedicated system packs

Categories:

Ref. No.
AVCMB – 10 | AVCMB – 20 | AVCMB – 50
Pack Size
10 ml | 20 ml | 50 ml
Presentation
1 x 8 / 1 x 2 ml | 2 x 8 / 2 x 2 ml | 5 x 8 / 5 x 2 ml
  • Reaction Type (Mode) : Kinetic
  • Reaction Direction : Increasing
  • Wave Length : 340nm
  • Flow Cell Temp. : 37°C
  • Zero Setting with : Distilled Water
  • Delay time : 300 seconds
  • Measuring Time : 180 Seconds
  • Reagent Volume : 0.5 ml ( 400 µl R1+ 100 µl R2)
  • Sample Volume : 25 µl
  • Factor : 8655
  • Linearity : 1000
  • Units : IU/L
  • High Normal : 24

Related Products