LiquiMAX CK-NAC

(IFCC Method)

INTENDED USE: Kit is use for the determination of CKNAC Iso Enzyme in Human Serum and Plasma.

Product Fearures:

  • Liquid stable, ready-to-use two reagents (R1 = Buffer, R2 = Substrate)
  • Kinetic Factor: 3987
  • NADP analogues are used for better stability
  • Linearity up to 2000 IU/L
  • 6 minutes increasing kinetic reaction
  • Measuring wavelength: 340 nm
  • Specimens: Serum and Plasma
  • Available as multipurpose reagents and dedicated system packs

Categories:

  • Ref No.
  • AVNAC-10
  • AVNAC-20
  • AVNAC-50
  • Pack Size
  • 10 ml
  • 20 ml
  • 50 ml
  • Presentation
  • 1 × 8 / 1 × 2 ml
  • 2 × 8 / 2 × 2 ml
  • 5 × 8 / 5 × 2 ml
  • Reaction Type (Mode): Kinetic
  • Reaction Direction: Increasing
  • Wave Length: 340 nm
  • Flow Cell Temp.: 37°C
  • Zero Setting with: Distilled Water
  • Delay time: 120 seconds
  • Measuring Time: 240 seconds
  • Reagent Volume: 0.5 ml (400 µl R1 + 100 µl R2)
  • Sample Volume: 25 µl
  • Factor: 3987
  • Linearity: 2000
  • Units: IU/L
  • High Normal: 190 (Adult Male)
  • 167 (Adult Female)